Fluorescein angiography

From Self-sufficiency
Jump to: navigation, search

Fluorescein angiography, or fluorescent angiography, is a technique for examining the circulation of the retina using the dye tracing method. It involves injection of sodium fluorescein[1] into the systemic circulation, and then an angiogram is obtained by photographing the fluorescence emitted after illumination of the retina with blue light at a wavelength of 490 nanometers. The fluorescein dye also reappears 12-24 hours in the patient urine, causing a yellow-green appearance.

Fluorescein angiography is one of several health care applications of this dye, all of which have a risk of severe adverse effects. See fluorescein safety in health care applications.

Equipment

  • Exciter filter: Allows only blue light to illuminate the retina. Depending on the specific filter, the excitation wavelength hitting the retina will be between 465-490 nm. Most only allow light through at a wavelength of 490 nm.
  • Barrier filter: Allows only yellow-green light (from the fluorescence) to reach the camera. Both filters are interference bandpass filters, which means they block out all light except that at a specific wavelength. The barrier filter only allows light with a wavelength of 525 nm, but depending on the filter it can be anywhere from 520-530 nm.
  • Fundus Camera with camera body containing black and white, or slide positive film. Also digital cameras tethered to computers have come into use since the late 1990s and are beginning to dominate the market today.

Technique

File:Fluorescein angiography.jpg
Retina during the effects of fluorescein angiography
  • Baseline color and black and white red-free filtered images are taken prior to injection. The black and white images are filtered red-free ( a green filter ) to increase contrast and often gives a better image of the fundus than the color image.
  • A 6 second bolus Injection of 2-5cc of sodium fluorescein into a vein in the arm or hand.
  • A series of black-and-white or digital photographs are taken of the retina before and after the fluorescein reaches the retinal circulation (approximately 10 seconds after injection). The early images allow for the recognition of autofluorescence of the retinal tissues. Photos are taken approximately once every second for about 20 seconds, then less often. A delayed image is obtained at 5 and 10 minutes. Some doctors like to see a 15 minute image as well.
  • A filter is placed in the camera so only the fluorescent, yellow-green light (530 nm) is recorded. The camera may however pick up signals from pseudofluorescence or autofluorescence. In pseudofluorescence, non-fluorescent light is imaged. This occurs when blue light reflected from the retina passes through the filter. This is generally a problem with older filters, and annual replacement of these filters is recommended. In autofluorescence, fluorescence from the eye occurs without injection of the dye. This may be seen with optic nerve head drusen, astrocytic hamartoma, or calcific scarring.
  • Black-and-white photos give better contrast than color photos, which aren't necessary since only one color is being transmitted though the filter.

Normal circulatory filling

times are approximate

  • 0 seconds — injection of fluorescein
  • 9.5 sec — posterior ciliary arteries
  • 10 sec — choroidal flush (or "pre-arterial phase")
  • 10-12 sec — retinal arterial stage
  • 13 sec — capillary transition stage
  • 14-15 sec — early venous stage (or "lamellar stage", "arterial-venous stage")
  • 16-17 sec — venous stage
  • 18-20 sec — late venous stage
  • 5 minutes — late staining

Fluorescein enters the ocular circulation from the internal carotid artery via the ophthalmic artery. The ophthalmic artery supplies the choroid via the short posterior ciliary arteries and the retina via the central retinal artery, however, the route to the choroid is typically less circuitous than the route to the retina. This accounts for the short delay between the "choroidal flush" and retinal filling.

Pathologic findings

Pathologic changes are recognized by the detection of either hyperfluorescence or hypofluroescence.

Causes of hyperfluorescence:

leaking defects (i.e. capillary leakage, aneurysm, neovascularization)
pooling defects
staining
transmission (filling) defects
abnormal vasculature

Causes of hypofluorescence:

blocking defect (i.e. blood)
filling defect (capillary blockage)

Among the common groups of ophthalmologic disease, fluorescein angiography can detect diabetic retinopathy, vein occlusions, retinal artery occlusions, edema of the optic disc, and tumors.

Other types of fluorescent angiography

Sources

  • Richard, Gisbert „Fluorescein and ICG Angiography“, THIEME 1998, ISBN 0-86577-712-8
  • Kanski, Jack „Clinical Ophthalmology“, 2006