Admittance and conductance in cardiac performance
This article may be too technical for most readers to understand. Please help improve this article to make it understandable to non-experts, without removing the technical details. The talk page may contain suggestions. (September 2010) |
This article may need to be wikified to meet Wikipedia's quality standards. Please help by adding relevant internal links, or by improving the article's layout. (August 2010) Click [show] on right for more details.
No reason has been cited for the Wikify tag on this article.
|
There is an established practice of using the electrical conductance of blood (PV loops) in heart ventricles to determine the instantaneous volume of the ventricle. This technique involves inserting a tetra-polar catheter into the ventricle and measuring conductance. This measured conductance is a combination of blood and muscle and various techniques are used to identify the blood conductance from the total measured conductance. Blood conductance can then be converted to volume using a linear (Baan) or a non-linear (Wei) relationship that relates conductance to volume.
Contents
Traditional conductance technology
This approach is based in the idea that the total conductance, G, of a fluid between two electrodes is a function of the fluid's conductivity (reciprocal of resistivity) and volume.
In cardiology, a tetra-polar catheter is inserted into the ventricle and a constant current is applied across the two outer electrodes. This generates an electrical field within the ventricle and the two inner electrodes measure a voltage generated due to the electric field. This measured voltage is proportional to conductance, which is related to blood volume (Baan assumes a linear relation whereas Wei uses a non-linear relation). When used in cardiology, the electric field generated is not limited to the blood (the fluid of interest) but also penetrates the heart wall, giving rise to additional conductance often called "parallel conductance" or "muscle conductance", Gm which must be removed.
Various techniques have been attempted to remove the Gm contribution with varying degrees of success. The most common method is the hypertonic saline technique which involves injecting a bolus of hypertonic saline into the ventricle to alter blood conductivity without affecting the surrounding muscle. Another less commonly used technique involves evacuating the ventricle of blood and measuring muscle conductance alone with a conductance catheter. Clearly both techniques are unreliable, somewhat invasive and fail to account for the continuous variation in Gm over the cardiac cycle.
Improved admittance technology
The Admittance technique is an improvement over the Conductance technique for the real-time removal of muscle conductance Gm. Blood and muscle respond very differently to alternating (AC) electrical currents. The fixed charges in muscle cells create a significant reactance that causes a phase shift (time delay) in the measured signal, relative to the excitation signal. Admittance technology uses this phase shift to determine the instantaneous muscle conductance and remove it from the total measured conductance.
Removal of Gm
The total Admittance (Y), of the blood filled ventricle is given by Y = Gb + Gm + iωCm where
- Gb is the measured conductance of the blood (the desired signal)
- Gm is the measured conductance of the cardiac muscle (unwanted signal)
- Cm is the measured capacitance of the cardiac muscle
- ω is the angular frequency of the excitation signal
The signals Gm and Cm are both properties of cardiac muscle and vary in a fixed ratio. Thus the ratio of Gm to Cm is equal to the ratio of muscle conductivity (σ) to muscle permittivity (ε). The ratio σ/ε is the constant of proportionality. Although both σ and ε are functions of the health of the heart tissue, they are relatively constant for short periods of time.
Using this proportionality, one can rewrite the equation for Gm as Gm = (σ/ε)Cm
Note that the imaginary component of Y depends only on the amount of muscle in the field of the catheter. This makes it easy to isolate by measuring the phase shift, φ, of the measured signal:
cos(φ) = (Gb + Gm)/Y
sin(φ) = ωCm/Y
Hence, Cm = Y.sin(φ)/ω
also, Gm = (σ/ε)Cm
Thus, blood conductance is determined as Gb = Y.cos(φ) - Gm
The desired conductance to volume conversion equation can be applied to this calculated blood conductance Gb to obtain blood volume.
Experimental results
Admittance technique involves the measurement of both phase angle and total conductance in the ventricle. Thus, it is possible to observe how the parallel conductance (muscle conductance) varies throughout the cardiac cycle. A plot showing both the blood and muscle contribution are shown in the figure. This plot of blood and muscle conductance shows how the parallel conductance due to muscle varies over the cardiac cycle.
References
- Chia-Ling Wei; Valvano JW; Feldman MD; Pearce JA (2005). "Nonlinear conductance-volume relationship for murine conductance catheter measurement system". IEEE Transactions on Biomedical Engineering. 52 (10): 1654–61. doi:10.1109/TBME.2005.856029.
- Chia-Ling Wei; Valvano JW; Feldman MD; Pearce JA (2007). "Volume Catheter Parallel Conductance Varies Between End-Systole and End-Diastole". IEEE Transactions on Biomedical Engineering. 54 (8): 1480–9. doi:10.1109/TBME.2007.890732.
- Krenz M (2009). "Conductance, admittance, and hypertonic saline: should we take ventricular volume measurements with a grain of salt?" (PDF). Journal of Applied Physiology. doi:10.1152/japplphysiol.01089.2009.
- Porterfield JE, Kottam ATG, Raghavan K, Escobedo D, Jenkins JT, Larson ER, Treviño RJ, Valvano JW, Pearce JA, Feldman MD (August 20, 2009). "Dynamic correction for parallel conductance Gp and gain factor α in invasive murine left ventricular volume measurements". Journal of Applied Physiology. 107 (6): 1693–1703. doi:10.1152/japplphysiol.91322.2008.
- Clarka JE, Kottamb A, Motterlinic R, Marber MS (March–April 2009). "Measuring left ventricular function in the normal, infarcted and CORM-3-preconditioned mouse heart using complex admittance-derived pressure volume loops". Journal of Pharmacological and Toxicological Methods. 59 (2): 94–9. doi:10.1016/j.vascn.2008.10.007.
- [1], Marc D. Feldman; Jonathan W. Valvano; John A. Pearce; Chia-Ling Wei, "Method and Apparatus for Determining Cardiac Performance in a Patient with a Conductance Catheter" (patent application)
- Wikipedia articles that are too technical from September 2010
- Articles with invalid date parameter in template
- All articles that are too technical
- Articles needing expert attention from September 2010
- All articles needing expert attention
- Pages with broken file links
- Articles that need to be wikified from August 2010
- All articles that need to be wikified
- CS1 maint: Multiple names: authors list
- 2Fix
- Cardiology